Thermo Fisher Scientific CyberScan DO 1500 User Manual

Page 13

background image

Instruction Manual

CyberScan DO 1500

10

3. If the unit has not been standardised, a series of dashes will

appear on the display rather than a number.


4. Press enter key to confirm selection and to go to next setup

page, OR press std to exit from this page without confirming the
change mode.


)

If you access the setup page from %Sat mode and if you have
calibrated one point (100% Sat) then you can view the last calibrated
Cal Factor. Should you have calibrated 2 points (100% and 0%sat)
then you can view last calibrated Cal Factor and 0% offset.

If you access the setup page from mg/L mode, and you have
calibrated mg/L then you can view the last calibrated value in mg/L.