14.3 Standard pH Buffers

The following table shows the various pH values at different temperature of the
solution during calibration.

Temperature

(ºC)

pH 1.68

(oxalate)

pH 4.01

(phthalate)

pH 7.00

(neutral

Phosphate)

pH 10.01

(carbonate)

pH 12.45

(Saturated

Calcium

Hydroxide

Solution)

0 1.67 4.01 7.12 10.32 13.43
5 1.67 4.01 7.09 10.25 13.21

10 1.67 4.00 7.06 10.18 13.00
15 1.67 4.00 7.04 10.12 12.81
20 1.68 4.00 7.02 10.06 12.63
25 1.68 4.01 7.00 10.01 12.45
30 1.69 4.01 6.99 9.97 12.29
35 1.69 4.02 6.98 9.93 12.13
40 1.70 4.03 6.97 9.89 11.99
45 1.70 4.04 6.97 9.86 11.84
50 1.71 4.06 6.97 9.83 11.70
55

- 4.08 6.97 9.81 -

60

- 4.10 6.98 9.79 -

70

- 4.12 6.99 9.76 -

80

- 4.16 7.00 9.74 -

90

- 4.20 7.02 9.73 -