9.2.4 P1.4 Select °C or °F Temperature Units

Available in pH 110 meter only

The pH 110 meter lets you select between °C and °F
units for temperature readings.

From measurement mode

1. Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P1.0.

3. Press the HOLD/ENTER key four times to

select parameter 1.4.

4. Press the MI/▲ or MR/▼ keys to toggle

between °C and °F.

5. Press the HOLD/ENTER key to confirm

selection and to return to the subgroup
menu. Press the CAL/MEAS key to return to
measurement mode.

See Figure 33.

°F

°C

SETUP

f

p 1.4

SETUP

C

p 1.4

MR

MI

Figure 33: P1.4 - Select