P2.0: Viewing Previous Calibration Data

This mode lets you recall previous calibration data,
which helps you know when to re-calibrate your meter.
This is a “view only” mode.

From measurement mode:

1. Press the SETUP key to enter Set up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P2.0.

3. Press the HOLD/ENTER key repeatedly to

view previous calibration data. See Figure 34.

4. When you have scrolled through all calibration

data, you will automatically return to the
subgroup menu. Press CAL/MEAS key to
return to measurement mode.

NOTES:

If there is no previous calibration data at a particular point, the
primary display will show “ ----“.