7. TROUBLESHOOTING

Problem Cause

Solution

No display

Batteries are not installed,
were improperly installed, or
are too weak

Install batteries with correct + / -
polarity. Replace with new
batteries

“LO” displays in
the LCD

Low battery

Replace batteries

Unstable
readings

a) Air bubbles in probe
b) Dirty

probe

c) Probe not immersed

enough

d) External noise pickup

or induction caused by
nearby electric motor

e) Broken

probe

a)

Tap probe to remove bubbles

b)

Clean probe & recalibrate

c)

Make sure sample entirely
covers the probe sensors

d)

Move or switch off interfering
motor

e) Replace

probe

Slow response

Dirty / Oily probe

Clean & recondition probe

Inaccurate
readings / can’t
calibrate

Probe guard not installed or
calibration solution incorrect

Install probe guard & replace
calibration solutions

Er1
COND 6+
TDS 6+

Attempted calibration value
was not within the ±40%
auto calibration window

Check the value of the conductivity
calibration solution.
Switch to manual calibration mode
and re-calibrate

Er1
SALT 6+

Salinity calibration error
CAL key was pressed in
Under range “Ur” or Over
range “Or” condition

Immerse the probe in a calibration
solution between 1.0 - 50 ppt

Er2
Temperature
calibration error

Auto calibration was not
within (0 – 50 °C)
temperature range

Ensure that the temperature is
within the acceptable range