Measurement, 1 with automatic temperature compensation (atc), 2 without atc (manual temperature compensation) – Thermo Fisher Scientific Eutech COND/TDS/Salt 6 Plus User Manual
Page 20

Instruction Manual
COND 6+, TDS 6+, SALT 6+
-15-
4. MEASUREMENT
Your meter is capable of taking measurements that incorporate temperature
measurements automatically (most common) or using a temperature which you input
manually (rare).
4.1
With Automatic Temperature Compensation (ATC)
To compensate your reading using temperature values as measured by your electrode,
simply attach the phono plug of the electrode to the meter. The measured reading is
automatically compensated to the specified normalization temperature (either 20 °C or
25 °C). The rate of compensation which is applied is the Temperature Coefficient.
See Section 5.5 – Temperature Coefficient.
See Section 5.6 – Normalization Temperature.
4.2
Without ATC (Manual Temperature Compensation)
You can use manual temperature compensation after manually entering the
temperature value of your process into the meter. The meter will then compensate from
this fixed value to the normalization temperature. Any temperature can be used between
0 and 50 °C. The default value is 25 °C.
To activate manual temperature compensation, simply unplug the temperature sensor
from the meter (mini phono plug) and follow the steps 2 thru 6 of Temperature
Calibration.
See Section 3.6 – Temperature Calibration.