-13-
3.5 Manual
Calibration
In Manual Calibration mode you are not limited to the conductivity calibration standards
listed previously in Table 1. This example shows a manual calibration sequence using a
12.00 mS/cm conductivity calibration standard.
1. If necessary, press MODE key to select
conductivity mode.
2.
Rinse the probe thoroughly with de-ionized water or
a rinse solution, then rinse with a small amount of
calibration standard.
3.
Dip the probe into the calibration standard. Stir the
probe gently to create a homogeneous sample.
Allow time for the reading to stabilize.
4. Press
CAL to enter conductivity calibration mode.
The [CA] indicator will appear briefly, then a value
will appear flashing.
NOTE: To exit calibration without confirmation, press
CAL to return to measurement mode.
5.
When the value is stable, press ▲ or ▼ as needed
to adjust the value to match your calibration
standard.
6. Press
ENTER to confirm the adjusted value. [CO]
will appear briefly indicating that the calibration was
successful. The meter returns to measurement
mode.
7.
Repeat steps 1-6 as needed with additional
calibration standards.