DO NOT TIP THE TUBE SO THAT THE COLOURED END IS DOWN AS THIS WILL CAUSE A

SPILLAGE.

3. The unit has been designed to conform to international safety specifications and is fitted with a re-

settable over-temperature cut-out. If, for any reason, the temperature rises above a safe level for

the instrument, the over temperature protection cuts out the heater, but the tubes continue to

rotate. Check the reason for the cut-out; if there is no obvious cause, reset the cut-out once the

unit has cooled. The reset, on the rear of the unit, is operated by pressing the centre of the button

indicated on page 19.

USER MAINTENANCE

O ring seals on the FHB16 hybridisation tubes
The O ring seals on the large 80mm diameter hybridisation tubes must be checked regularly and

replaced before a leak occurs. Each tube is provided with 6 spare O rings and further O rings can be

obtained from Bibby Scientific Ltd. or your Techne dealer.

Hybridisation tubes
The hybridisation tubes must be checked regularly and replaced if any damage occurs. Any damage

to the ends of the tubes can cause them to leak or break when the end caps are inserted. Further

tubes can be obtained from Bibby Scientific Ltd. or your Techne dealer.

Cleaning your HB-1D
Before cleaning your unit ALWAYS disconnect from the power supply and allow to cool.

You can clean the case of the HB-1D with a cloth dampened in soapy water. No part of the unit

should be immersed in solvents. Do not use acetone or abrasive cleaners.

DO NOT USE ETHANOL OR METHYLATED SPIRITS ON THE ACRYLIC DOOR; these may make

the acrylic go cloudy.

Before using any cleaning or decontamination method except those recommended in this manual,

the responsible body should check with Bibby Scientific Ltd. that the proposed method will not

damage the equipment.

If you spill the fluid
Because of the nature of operation of the HB-1D, the complete inside must be cleaned if a spillage

does occur. This is to remove any airborne particles which may have been blown around the system

and to guarantee full operator safety.

Switch the HB-1D off and disconnect it from the power supply. Follow this sequence of operations

when the HB-1D has been allowed to cool below at least 50°C:

Remove all glassware from the HB-1D.

Remove the plastic drip tray.

Clean all surfaces of the inner chamber using a suitable cleaning and decontaminating

solution. Be careful of the drive shafts.


Cleaning and decontamination materials recommended by Bibby Scientific include:

Lipsol®

Decon® 90

Neutracon®

Other mild detergent