300 Gas Detector Manual – TM 794100, Rev. D, May 2004

16


Notes:
1 Four Gas types can be selected for maximum compatibility to Gas spectral

properties. The selection of the appropriate type is defined by pre-setting of
SW1-3 and SW1-4 according to the following table:

Table 8. Gas Type Calibration Setup

bit # 0 position

(SW1-3)

bit # 1 position

(SW1-4)

Detected Gas

Type

Gas type as

defined in

Section 2.3.6

off

off

Methane

Gas 1

on

off

Mixture

Gas 2

off

on

Propane

Gas 3

on

on

Mixture

Gas 4


2


In Model 301 the standard full scale is 2.5 LEL.m and 1 LEL.m by DIPSwitch.

Address Selection Switch

The Address selection Switch provides up to 64 different addresses for the
detector upon a network installation. Each detector can be recognized at its
specific address in RS485 network. The address should be determined by a pre-
setting of SW2-1 to SW2-6 switches.

Attention!

SW2-7 and SW2-8 should be always at on.