5

Temperature/Humidity Range

Operating:

Stated Accuracy:

18

°

C to 28

°

C

(64

°

F to 82

°

F)

Useful Range:

0

°

C to 40

°

C

(32

°

F to 104

°

F)

Storage:

40

°

C to 65

°

C

(

40

°

F to 149

°

F)

Humidity:

10% to 90%
(non-condensing)

Dimensions

3 cm D x 8.4 cm W x 15.8 cm H
(1.2 in x 3.3 in x 6.2 in)

Weight with batteries: 227 grams (8 ounces)

Ingress protection: Meets IEC-529 IP-54 for dust

and water resistant enclosures.

Intrinsic safety

This product is energy limited for intrinsically safe
operation in hydrogen atmospheres per Class I,
Division 1, Groups A, B, C, and D hazardous
(classified) locations for UL per UL913 and CSA
per C22.2 No. 0-M1982 and No. 157-M1987.
Maximum surface temperature: 135°C (T4); UL
file No. E182612 (1997).

Compliance (For CE Mark):

EN61326-1/A1: 1998 (EU EMC Directive)