5. Allow the reading to stabilize. If the displayed
temperature is higher or lower than the refer-
ence temperature, use the MAX
s
key to in-
crease the displayed reading or the
MIN
t
key to decrease the displayed reading
until the reference temperature is displayed.
The MIN
t
or MAX
s
key must be pressed
at least once. The CAL annunciator should
remain blinking during this procedure.
6. Press the HOLD
key to lock the T2 offset
calibration in, or press the CAL key to skip T2
offset calibration and proceed with T1 slope.
If T1 or T2 offset calibration is skipped, then the
slope calibration for that input will also be skipped.
7. Slope Calibration: Place the probe(s) at the
higher reference temperature.
8. The temperature is displayed on the main dis-
play and "Hi" is displayed on the lower left dis-
play.
9. Allow the reading to stabilize. If the displayed
temperature is higher or lower than the refer-
ence temperature, use the MAX
s
key to
increase the displayed reading or the
MIN
t
key to decrease the displayed reading
until the reference temperature is displayed.
The MIN
t
or MAX
s
key must be pressed
at least once. The CAL annunciator should
remain blinking during this procedure.
10. Press the HOLD
key to lock the T1 slope
calibration in and advance to T2 slope calibra-
tion, or press the CAL key to skip T1 slope
calibration. In either case you will now be in T2
slope calibration.