Pyromation Series 423 User Manual
Page 18
Page 17
= UP
= ENTER
= DOWN
{Step 1}
{Step 2}
{Step 3}
{Step 4}
17-1
CJC CALIBRATION PROCEDURES
Power panel meter for at least 30 minutes.
Enter menu program, go to the
CAL
mode. Press the down
button and hold for approximately 1.5 seconds until the meter
reads
SHRT
, then release the button. Failure to hold the button
for 1.5 seconds will terminate the calibration process.
Press the down button until
CJC
appears.
Now that
CJC
appears, place the thermometer across the
terminal strip and do not remove until the temperature stabi-
lizes. The reading from the thermometer is going to be the new
CJC
temperature. Press and release the enter button. Use the
up or down button to set the value, then press the enter button
three times until V3.1 appears. Calibration is now complete.
CJC COMPENSATION
Figure 29