1 password menu, 2 calibration menu – Pyromation Series 423 User Manual

Page 15

background image

14-1 PASSWORD MENU

The calibration routine is for thermocouples and RTD's only. Described below is

the menu program for calibration. Each calibration can be independently done

using the down arrow to scroll to the next routine. Push the up arrow

anywhere to exit the calibration routine.

Figure 24

Figure 23

14-2 CALIBRATION MENU

Described below is the menu for selecting menu passwords.

or

028

YES / NO

IF YES

ENTR

PWRD

TO PICK

NEW VALUE

RETURN WHEN

AT DESIRED

PASSWORD

NEW PWRD

PWRD

.
.

WAIT

V3.1

READING

WAIT

CAL

USING COPPER WIRE JUMPER TERMINALS 5,6,7

PUSH TO SKIP 50mV

50mV

DISCONNECT JUMPERS

INPUT 50.040mV THROUGH COPPER WIRE TO TERMINALS 5,6

PUSH TO SKIP 100 OHMS

WAIT

100 OHMS

DISCONNECT 50.040mV

INPUT 100.0W ACROSS TERMINALS 5,6,7

WAIT

PUSH TO SKIP CJC

PUSH TO SKIP SHORT

HOLD UNTIL METER INDICATES SHRT

(approximately 1.5 seconds)

Equipment Needed for Calibration

50.040 mV + .005 Power Source
100.03W + .01W Resistor

1

+ .1°F Thermometer

Note

1

: Adjust a potentiometer to achieve

100.03W

or

CJC

DISCONNECT 100W

TO VIEW

or

77.0

TO SET NEW VALUE

AGAINST THERMOMETER

PWRD

.

.

Page 14