16-1

RTD CALIBRATION PROCEDURES

Power panel meter for at least 30 minutes.

Enter menu program, go to the

CAL

mode. Press the down

button and hold for approximately 1.5 seconds until the meter

reads

SHRT

, then release the button. Failure to hold the button

for 1.5 seconds will terminate the calibration process.

Install jumpers between terminals 5, 6, and 7 using copper

wire. Press and release the enter button. Wait until 50mV

appears. Disconnect jumpers.

Press the down button to skip 50mV input.

Now that 100 ohm appears, install jumper between terminals

6 and 7, and insert the 100W resistor into terminals 5 and 6.

Then press and release the enter button. This simulates a 3

wire RTD. Wait until

CJC

appears. Disconnect jumper and

resistor.

Press the up button to save and exit calibration, then press the

enter button three times until V3.1 appears. Calibration is now

complete.

= UP

= ENTER

= DOWN

7

JUMPER

6

5

5

6

7

RTD

JUMPER

Figure 28

Figure 27

{Step 1}

{Step 2}

{Step 3}

{Step 4}

{Step 5}

{Step 6}