4 - 20mA CALIBRATION PROCEDURES
(All steps must be completed)
= ENTER
{Step 1}
{Step 2}
{Step 3}
{Step 4}
{Step 5}
{Step 6}
{Step 7}
{Step 8}
{Step 9}
{Step 10}
Power panel meter for at least 30 minutes.
Enter menu program, go to the
SNSR
menu. Press the
ENTER
button.
MAMP
should appear on the display. If not, press the
down button until
MAMP
is displayed. Press the
ENTER
button.
Select proper resolution for the decimal point by pressing the
down button. When proper resolution has been selected
press the
ENTER
button.
Select proper scale indication by pressing the down button.
If °F or °C is not desired, select
OFF
. Press the
ENTER
button.
(Note: If
OFF
was selected, scale indication stickers have
been supplied for the front display)
Now that
ZERO
appears, press the down or up button, then
select the desired display reading.
Once the display reading has been established, connect a
4mA source to the input of the meter. Refer to Figure 16 for
proper connections. Press the
ENTER
button. Wait until
SPAN
appears.
Now that
SPAN
appears, press the down or up button, then
select the desired display reading.
Once the display reading has been established, connect a
20mA source to the input of the meter. Refer to Figure 16 for
proper connection. Press the
ENTER
button,
WAIT
until
UNDR
appears.
Now that
UNDR
appears, press the down or up button, then
select the desired display reading for the under indication.
Press the
ENTER
button.
Now that
OVER
appears, press the down or up button, then
select the desired display reading for the over indication.
Press the
ENTER
button. Calibration is now complete.
Refer to 4 - 20mA menu on page 12 for the above steps (fig. 22)