27

Example:

Level 100 %:

4000 mm

Density:

1 g/cm3

Density correction:

0.9 g/cm3

Span end point:

The span (end-point) must be re-calibrated (with or without pressure) to 4000 mm in
order to prevent a 4000 mm level tank from being overfilled.

6.5.5

Main Menu: Language

A change or correction in the density causes a change in the span end-
point’s unit of measure (mm, m, inch, feet). The span end-point must be
re-calibrated when changing the medium to be measured (due to a
change in density).

4000 mm 1 g/cm

3

0.9 g/cm

3

-------------------------------------------------- 4444

mm

=