Martel Electronics 320 User Manual
Page 40

The LPS battery voltage should be between 10.6 and
9.2 volts. When the voltage falls below 8.8 volts, the
LPS will automatically turn itself off to prevent
damage to the batteries due to possible cell reversal.
The battery charger used for the Model 320 should be
used for the LPS.
APPLICATION NOTES
USING
THE
MODEL
320
PRESSURE
CALIBRATOR WITH A HAND PUMP
1. Safely take the instrument to be tested out of
service, equalize it, and vent it.
2. Vent the low side to atmospere and connect the
high side as shown, being very careful that there are
no leaks. Note: It is best to drain the high and low
sides to prevent liquids from entering the calibration
tubing.
3. Connect the test leads to the Model 320 mA input
and to the instrument to be calibrated to form a series
loop.
38
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)