Martel Electronics 320 User Manual
Page 26

Select “2” if the pressure input to the pneumatic
transmitter is being measured on the right port of the
Model 320.
After the percent function setup is completed, the
function is activated by pressing the % key.
4 = H
2
Oref (H
2
O Reference Temperature)
The H
2
O Reference refers to the temperature
correction for inches and centimeters of water
measurements. There are three choices: 20
°
C, (ISA
Standard); 4
°
C, (Scientific Standard); and 60
°
F,
(American Gas Association Standard). Press the
number for the desired temperature correction.
5 = Owner (Change Owner’s Name)
The Model 320 calibrator contains a security feature
that allows the owner to enter a name or other
24
1 = 20C 2 = 4C
3 = 60F
See also other documents in the category Martel Electronics Equipment:
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)