Martel Electronics 320 User Manual
Page 17

Once the proper zero and full scale values have been
entered, the operator presses the % key to activate the
percent function. The zero and span adjustments to
the transmitter can now be quickly adjusted by
applying aproximately 0% and 100% inputs to the
transmitter as shown on the display, and adjusting the
output for 0% error as shown on the display.
After the zero and span adjustments have been made,
the linearity can be quickly checked by applying
various intermediate inputs and noting the error
directly in percent. If there is no error, 0% will be
displayed for any pressure input.
The application note on page 45 gives examples for
using the percent function.
Pressing the % key a second time will return the
display to the selected engineering units. The percent
mode is also terminated if a new function or
engineering unit is selected.
TRIP RESET KEY
The Trip Rst key is used in conjunction with the trip
15
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)