beautypg.com

Caution – Martel Electronics MC1000 User Manual

Page 7

background image

5

Caution

To avoid possible damage to calibrator or

to equipment under test:

• Disconnect the power and discharge all high-

voltage capacitors before testing resistance
or continuity.

• Use the proper jacks, function, and range for

your measurement or sourcing application.

• To avoid mechanically damaging the pressure

module, never apply more than 10 ft-lb. of
torque between the pressure module fittings,
or between the fittings an the body of the
module.

• To avoid damaging the pressure module from

overpressure, never apply pressure above the
rated maximum printed on the module.

• To avoid damaging the pressure module from

corrosion, use it only with specified materials.
Refer to the pressure module documentation
for material compatibility.