beautypg.com

S" commands select sensor type – Martel Electronics MC1000 User Manual

Page 20

background image

Serial Input

Description

A

mA measurement

a

mA source

I

mA loop

i

mA 2W Sim

V

Volts measurement

v

Volts source

M

mV measurement

m

mV source

E

HVAC measure

e

HVDC measure

K

Khz measurement

k

Khz source

H

Hz measurement

h

Hz source

P

CPM measurement

p

CPM source

W

2-wire measurement (Ohms and RTDs)

X

3-wire measurement (Ohms and RTDs)

Y

4-wire measurement (Ohms and RTDs)

T

Thermocouple measurement (default Type J) use
"S" command to select sensor type

t

Thermocouple source (default Type J) use "S"
command to select sensor type

C

Selects Centigrade ( T/C - RTD )

F

Selects Fahrenheit ( T/C - RTD )

R

RTD measurement mode (default Pt100 385) use
"S" command to select sensor type

r

RTD measurement mode (default Pt100 385) use
"S" command to select sensor type

0-9

Enter a source value using ascii characters

-,.

0,1,2,..,9,-,.terminated by (carriage

return)

D

Single Broadcast of most recent display value and
units

d

Single Broadcast of most recent display value

"S" Commands Select
Sensor Type

Serial Input

Selection Entry

Description

S

No.

T/C Type

RTD Type

1

J

Pt100 (3926)

2

K

Pt100 (385)

3

T

Pt100 (3916)

4

E

Cu10

5

R

Pt200 (385)

18