Command
Description
PULSE_CNT?
Returns the number of pulses in the pulse train
REMOTE
Puts the calibrator in remote mode
SIM
Sets the output for mA simulation
SIM?
Returns the output of the mA simulation
TRIG
Starts and stops the pulse train in pulse mode
TRIG?
Returns TRIGGERED when a pulse train is on. Returns UNTRIGGERED when the pulse
train is off.
VAL?
Returns the measured values
V_RANGE
Sets voltage range
V_RANGE?
Returns voltage range
Table 7: Parameter units
Units
Meaning
MA
milliamps of current
MV
Voltage in millivolts
V
Voltage in volts
CPM
Frequency in cycles per minute
Hz
Frequency in Hertz
KHz
Frequency in kiloHertz
Table 8: Error codes
Error Number
Error Description
100
A non-numeric entry was received where it should be a numeric entry
101
Too many significant digits entered
102
Invalid units or parameter value received
103
Entry is above the upper limit of the allowable range
104
Entry is below the lower limit of the allowable range
105
A required command parameter was missing
110
An unknown command was received
112
The serial input buffer overflowed
113
Too many entries in the command line
114
The serial output buffer overflowed
115
Output is overloaded
116
Calibrator not in pulse train mode when TRIG was received
117
An invalid FREQ_TYPE was received