Auto-test, Store / recall a measured result – KYORITSU 6050 User Manual
Page 27
̶ 25 ̶
7.2.3 Testing
(1) Press the TEST BUTTON.
(2) Measured result is displayed on LCD.
If the measured result is 100V or more, "UcH V" is displayed on
LCD.
Note: When Test Lead removed during a measurement, "no" is
displayed on LCD and a measurement will be stopped. Please
check Test Lead is rightly connected.
If "I⊿n" setting is grater than the rated residual current of RCD,
RCD will trip and "no" may be displayed on LCD.
8. AUTO-TEST
When applying voltage to the instrument, with Test button is turned and
being locked, voltage value is displayed on LCD. And 3sec. later, a
measurement on selected function is performed automatically. Be sure
to release the locked Test button after a measurement.
9. STORE / RECALL A MEASURED RESULT
Measured result at each function can be stored in the memory of the
instrument. (MAX : 300)
9.1 How to store data
Store the result according to following sequence.
(1) After a measurement, press MEMORY MODE SWITCH ("MEM
MODE" SWITCH) while measured result is being displayed. Then
instrument becomes MEMORY MODE, and "
MEM
" mark appears
on the LCD. The action of each switch changes as shown under
each switch.
(2) Select the Data No you want to store with MEMORY SELECT
SWITCH ("MEM ▼▲ "SWITCH).
(3) Decide it with ENTER SWITCH("ENT" SWITCH).
(4) Select the Place No you want to store with MEMORY SELECT
SWITCH ("MEM ▼▲ "SWITCH).