KYORITSU 6050 User Manual
Page 14
̶ 12 ̶
Function
Operating error of trip current
Ч1/2
-10%∼0%
Ч1
0%∼+10%
Ч5
0%∼+10%
Auto Ramp
-10%∼+10%
● Operating Errors of trip current (IEC 61557-6)
The influencing variations used for calculating the operating error are
denoted as follows:
Temperature : 0℃ and 40℃
Earth electrode resistance :
I n(mA)
10
30
100
300
500
1000
Earth electrode resistance(Ω)
UL50V
2000
600
200
130
80
40
UL25V
2000
600
200
65
40
20
table.1
System voltage : 230V+10%-15%
Supply voltage : 8V to 13.8V
#
CAUTION
Even though the test current in the 2000 ohm range (15mA test
current) is low some RCD's may trip due to sensitivity or where
there may already be additional leakage in the circuit being tested.
The Loop impedance in a TN system is small and therefore it is
not recommended to test in the 2000 ohm range. RCD's will have
to be bridged to avoid tripping when using other test ranges.
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)