KYORITSU 6050 User Manual
Page 12
̶ 10 ̶
Function
Voltage
Measurement
Range
100-260V
L-L corresponding range:
100-440V
Rated Voltage
(AC)
100-260V
L-L corresponding range:
100-440V
Accuracy
±(2%rdg+4dgt)
RCD
Rated Voltage
(AC)
L-PE :
+10%
230V
50Hz
-15%
Accuracy
± (1%+3dgt)
Function
Trip Current
Trip Time
Ч1/2
Ч1
Ч5
DC
Auto Ramp
-8%∼-2%
+2%∼+8%
± 10%
± 4%
RCD Trip Current (I⊿n) and Trip Current Duration
Note: Bias current (DC6mA) added to DCrange.
● Instrument dimensions
● Instrument weight
186Ч167Ч89mm
980g
RCD Trip
Current
Duration
(ms)
RCD Trip Current I⊿n(mA)
Ч 1/2
Ч 1
Ч 5
DC
Auto
Ramp
10
1000
1000
200
1000
30
1000
1000
200
1000
100
1000
1000
200
1000
300
1000
1000
n.a.
1000
500
1000
1000
n.a.
200
1000
1000
200
n.a.
n.a.
n.a.
Goes up by 10% from 20% to110% of I⊿n.
300ms×10
n.a. :not applicable
See also other documents in the category KYORITSU Tools:
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)