5 ac voltage ranges and accuracy, 6 input impedance, 7 battery life – KYORITSU 6017 User Manual
Page 7

5
3-4 Earth Resistance Ranges and Accuracy: Simplified Measurement
Function
Measuring Range
Accuracy
Output current is the same as in precision measurement.
X 1 Ω
0-12 Ω
0-120 Ω
0-1200 Ω
X 10 Ω
X 100 Ω
± 3% of full scale value
(Value subtracted test lead resistance)
3-5 AC Voltage Ranges and Accuracy
Function
Measuring Range
Frequency
Accuracy
600V AC
0-600V
50/60Hz
± 3% of full scale value
Earth Voltage: 60V AC
0-60V
3-6 Input Impedance
Function
Input Impedance
2.8MΩ
135kΩ
600V AC
Earth Voltage: 60V AC
3-7 Battery Life
Number of measurements within effective battery voltage range.
(5-second measurement /25-second pause cycle)
Function
Test Resistance
Number of Measurements
More than 1000 times
More than 2000 times
More than 3000 times
More than 3500 times
More than 3000 times
More than 4000 times
1MΩ
0.5MΩ
0.25MΩ
0.125MΩ
1200Ω
1200Ω
Earth ( 3 Pole )
Earth ( 2 Pole )
X 100 Ω
X 100 Ω
125V
250V
500V
1000V
Insulation Resistance