3. Specifications
3-1 Standards
Designed to meet:
IEC 61557-2: Insulation Tester
IEC 61557-5: Earth Resistance Tester
IEC 61010-1 over-voltage category CAT. e 600V, pollution degree 2
3-2 Insulation Resistance Ranges and Accuracy
Function
1000V/2000MΩ
1000V
0-2000MΩ
50MΩ
1000VDC+20%
– 0%
1-1.6mADC or less
(Measured with an ammeter with a voltage drop of 500mV or less)
1-1.2mADC or less
(Measured with an ammeter with a voltage drop of 500mV or less)
at 1MΩ
2-1000MΩ
± 5% of indicated value
Ranges other than the above range, 0 and
∞
± 10% of indicated value
0.7% of scale length
0.1-50MΩ
0.05-20MΩ
0.02-10MΩ
at 0.5MΩ
at 0.25MΩ
at 0.125MΩ
500VDC+20%
– 0%
250VDC+20%
– 0%
125VDC+20%
– 0%
2MΩ
1MΩ
0.5MΩ
0-100MΩ
0-50MΩ
0-20MΩ
500V
250V
125V
500V/100MΩ
250V/50MΩ
125V/20MΩ
Rated Test Voltage
Measuring Range
Mid-scale Value
Output Voltage at Open
Circuit
Short-circuit Current
Rated Test Current
Accuracy in Primary Effective
Measuring Range
Accuracy in Secondary
Effective Measuring Ranges
Accuracy at 0 and ∞
3-3 Earth Resistance Ranges and Accuracy: Precision Measurement
Function
Measuring Range
Accuracy
Output Current
X 1 Ω
0-12Ω
± 3% of full scale value
(The earth resistance at auxiliary earth spike is
within 10kΩ)
2.5mA AC max., 820 ±10Hz
0-120Ω
0-1200Ω
X 10 Ω
X 100 Ω