5-2 Mechanical Zero Adjustment

With the function selector switch set to the OFF position and without pressing the
test button, turn the zero adjuster with a screw driver so that the pointer lines up
with the ∞ mark on an insulation resistance scale.

5-3 Connecting Test Leads

Slide the sliding shutter over the terminals for precision earth resistance
measurement to reveal the probe socket. Insert the test probe connector into the
probe socket correctly as shown below.

The instrument is shipped with the slide shutter over the terminals for precision
earth resistance measurement. Use these terminals only for precision earth
resistance measurement.

DANGER

¡

When the test button or the remote control switch is pressed with the
function selector switch set to an insulation or earth resistance range, do
not touch the tips of the test probes, where a high voltage is present.

Probe Socket

Test probe connector