Technical data – KROHNE LS 6500 EN User Manual
Page 28
7
TECHNICAL DATA
28
LS 6500
www.krohne.com
03/2013 - 4000649103 - MA LS 6500 R03 en
Materials
Sensor housing
Stainless Steel 1.4301 / 304
Process connection
Stainless Steel 1.4404 / 316L
Sensor insulation
Virgin PEEK, FDA conform
Electrical connection
Cable gland M16: Plast or Nickel-plated brass
Plug M12: Nickel-plated brass
Process connections
Standard
Hygienic G½; DN38
Other
For other hygienic process connections, e. g. Tri-clamp, 11851, Varivent see
data sheet "Accessories".
Electrical connections
Power supply
Non-Ex / Ex nA: 12…36 VDC, 70 mA max.
Ex: 12…30 VDC, 70 mA max.
Power consumption
1.7 W
Power-up time
< 2 s
Reaction time
Max. 0.1 s
Damping
0...10 s
Cable entry
M16 cable gland or M12 (4 pole Lumberg)
Output
Output (active)
Max. 50 mA, short-circuit and high-temperature protected
Output type
PNP, NPN or digital output
Output polarity
See drawing in chapter "Electrical connection".
Active "Low"
NPN and digital output; (-VDC + 2.5 V) ± 0.5 V; R
load
= 1 kOhm
Active "High"
PNP and digital output; (VDC - 2.5 V) ± 0.5 V; R
load
= 1 kOhm
Factory settings
Measure: ε
r
> 2; damping: 0.1 s
Approvals and certifications
CE
This device fulfils the statutory requirements of the EC directives.
The manufacturer certifies successful testing of the product by applying the
CE marking.
ATEX
ATEX II 1G Ex ia IIC T5 (pending)
ATEX II 3G Ex nA II T 5 (pending)
ATEX II 1D Ex tD A20 IP67 100°C
Other standards and approvals
Other standards and approvals
Other standards and approvals
Other standards and approvals
Electromagnetic compatibility (EMC)
EN 61326
Vibration resistance
IEC 68-2-6, GL test 2 (standard and DN38 connection)
Hygiene
3A for G½ and DN38, FDA conform materials
.book Page 28 Tuesday, March 12, 2013 11:33 AM