ELECTRICAL CONNECTIONS
18
LS 6500
www.krohne.com
03/2013 - 4000649103 - MA LS 6500 R03 en
4.5 Configuration tool
The configuration tool can be ordered optionally to configure the LS 6500 switches.
Scope of delivery:
• Interface unit
• CD with software and product drivers (DTM)
• USB cable
• Cable with 2 alligator clips
The configuration tool connects the LS 6500 with a computer. With the corresponding software,
it is possible to get an online communication with the LS 6500.
By using this tool, device information like serial number, switching point range and tag numbers
are displayed on the computer. Settings as switching point, damping, polarity and output can be
changed.
Self-learning function for the contact output is possible as well as reset function to the default
values of the switching point.
By fine-tuning of the switching point settings, LS 6500 could differentiate between various
products which are covering the sensor. In other words, the device could be set to trigger on a
specific product and ignore a second product. Basis for this would be a different ε
r
value of these
two products.
Connection type
T
amb
Max. medium temperature Note
[°C]
[°F]
[°C]
[°F]
Standard and DN38
connection
-40...+85
-40…+185
+85
+185
-
-40...+60
-40...+140
+95
+203
1
-40...+40
-40...+104
+115
+239
1
Sliding connection
100 mm / 3.9"
-40...+85
-40…+185
+85
+185
-
-40...+60
-40...+140
+150
+302
1
-40...+40
-40...+104
+175
+347
1
Sliding connection
250 mm / 9.8"
-40...+85
-40…+185
+85
+185
-
-40...+60
-40...+140
+195
383
1
-40...+40
-40...+104
+200
+392
1, 2
1 If the sensor tip at the device is the only part in contact with the media.
2 Max. allowed medium temperature.
.book Page 18 Tuesday, March 12, 2013 11:33 AM