3.5 Installation of sliding connection

The following drawing shows how the sliding connection can be used for at least 4 applications:

1 Mounted at the top of a tank to adjust to a certain level.

2 Serving as a cooling neck in high media temperature applications.

3 Adjusted to place the sensor tip deeper inside the tank (for lumpy or sticky media).

4 To reach in through insulation material.

Figure 3-2: Possible applications for sliding connection

CAUTION!
The LS 6500 with sliding connection can be mounted with a static pressure up to 16 bar / 232 psi.
To prevent personnel injuries, it is essential that the safety chain is mounted correctly and
undamaged.