Technical data, 2 technical data – KROHNE LS 6500 EN User Manual
Page 27
TECHNICAL DATA
7
27
LS 6500
www.krohne.com
03/2013 - 4000649103 - MA LS 6500 R03 en
7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Electromagnetic wave, 100..180 MHz
Application range
Level detection, dry-run protection and media separation of liquids and
solids.
Design
Construction
The measurement system consists of a measuring sensor and the electronic
unit which is available in a compact version. The switching point is signalled
by a blue LED indication through the housing cover.
Options
Sliding connection / extension for high-temperature applications
Teach-In function for applications where the medium is hard to detect.
Accessories
Comprehensive range of adapters and process connections for hygienic
installation. Please refer to the specific data sheet "Accessories".
Measuring accuracy
Resolution
±
1 mm / ±0.04"
Hysteresis
±
1 mm / ±0.04"
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Temperature
+20°C ±5°C / +70°F ±10°F
Pressure
1013 mbar abs. ±20 mbar / 14.69 psig ±0.29 psig
Relative air humidity
60% ±15%
Operating conditions
Temperature
Temperature
Temperature
Temperature
Ambient temperature (T
amb
)
-40…+85°C / -40…+185°F
Process temperature
-40…+85°C / -40…+185°F (short version and DN38 connection)
< 1 hour, T
amb
< 60°C / 140°F: -40…+140°C / -40…+284°F
-40 …+200°C / -40...+392°F (with sliding connection)
Pressure
Pressure
Pressure
Pressure
Ambient pressure
Atmosphere
Process pressure
Standard and DN38 connection: max. 40 bar / 580 psi
Sliding connection: max. 16 bar / 232 psi
Other conditions
Other conditions
Other conditions
Other conditions
Protection category
(acc. to EN 60529)
IP67 equivalent to NEMA 4X
Installation conditions
Installation
In any position. For detailed information refer to chapter "Installation".
Dimensions and weights
For detailed information refer to chapter "Dimensions and weights".
.book Page 27 Tuesday, March 12, 2013 11:33 AM