Operation

5.1 Start-up

Before connecting to power, please check that the system has been correctly installed.
This includes:

• The device must be mechanically safe and mounted in compliance with the regulations.
• Check the leak-tightness at the sleeve.
• Make sure that the cable gland is tight or, as the case may be, the M12 plug is properly

screwed down.

• The power connections must have been made in compliance with the regulations.
• Check that the electrical operating data of the power supply are correct.

• Switching on the power.

i

Check for correct switching function.

5.2 Fault diagnosis and corrective action

Fault

Cause

Action / elimination

LED not "on"

Sensor cap not in contact with

liquid product

-

Supply voltage < 12 V, permittivity

too low

Measure voltage at Pin 1 and 2

(M12: pin 1 and 3)

No switching output

Cable break

Check continuity of cables

Incorrect polarity of supply

Reverse terminals 1 and 2 (M12:

terminal 1 and 3)

Short-circuit

Check wiring