beautypg.com

Technical data – KROHNE OPTIBAR P 3050 C EN User Manual

Page 37

background image

TECHNICAL DATA

7

37

OPTIBAR P 3050 C

www.krohne.com

03/2012 - 4001953401 - MA OPTIBAR P3050C R02 en

Operating conditions

Temperature limits

Temperature limits

Temperature limits

Temperature limits

Operating temperature

Blind version: -40…+85°C / -40...+185°F

With LCD graphic display: -20…+70°C / -4…+158°F

Ambient temperatures below -10°C / +14°F may affect the readability of the

display.

Process temperature

-40…+85°C / -40...+185°F

Storage temperature

-20…+70°C / -4…+158°F

Other conditions

Other conditions

Other conditions

Other conditions

Protection category acc. to IEC 529 /

EN 60529

Standard: IP65

Optional: IP67 with internally ventilated cable for gauge pressure sensors

Installation conditions

Installation

Can be installed in any position, zero point or position correction may be

required following installation.

Maximum error through mounting position: <3.5 mbar / <0.05 psi

Dimensions and weights

For detailed information refer to section "Dimensions and weights".

Materials

Wetted parts

Stainless steel W.1.4404 (AISI 316L)

Non-wetted parts

Stainless steel W.1.4404 (AISI 316L)

Internal housing cover gasket: EPDM

Version with display: Makrolon

®

Process connections

Standard

G½-B acc. to DIN EN 837-1

NPT versions

½

"-14 NPT - female

½

"-14 NPT - male

Process connections with flush

mounted diaphragm

In preparation

Electrical connection

Supply voltage

12...45 VDC

Output signal

4...20 mA, 2-wire

Damping

0.1 s

Max. load resistance (current output) R

Load

[KΩ] = (U

B

[V] - 12 V) / alarm current max. [mA]

with U

B

= supply voltage

Initialisation time

10 s

Alarm current

Configurable as high alarm (21 mA) and low alarm (3.6 mA) using optional LCD

display

Cable feedthroughs

M16 in plastic, nickel-plated brass or 316L stainless steel

.book Page 37 Thursday, March 8, 2012 9:29 AM