Page 13: Installation and operating instructions
DWM 1000
ERRATA
03/2012
– Errata– MA DWM 1000_2000 R01 en
www.krohne.com
Electromagnetic
Flow Switch
DWM 1000
ERRATA
INFORMATION!
The data given on this errata sheet is applicable to DWM 1000 only.
Installation and Operating Instructions
For additional safety instructions about DWM 1000 devices, refer to page 5
Use one safety device per fuse (In = 100mA) in the power supply circuit. Use a switchgear device
(switch, circuit-breaker) to disconnect the signal converter adjacent to the device. Switchgear must
fully isolate the device (disconnect the 2 power lines), agree with the IEC 60947-1 and IEC 60947-3
standards and be identified as switchgear for this device.
Make sure that the electrical cable diameter is between 8 and 12 mm. Make sure that the cable gland
holds the cable tightly. Electrical cables must have a temperature rating that agrees with the operating
conditions.
Use a damp cloth to clean the device. Do not use aggressive or corrosive cleaning agents.
Do not open the device when it is energized. There is a risk of injury or death from dangerous
voltages if you touch internal components after the device is connected to power supply. Do
not operate the device without its cover.
De-energize the device and disconnect the power supply before you do work on the electrical
connections. Refer to the voltage data on the nameplate!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
Obey the national regulations for electrical installations!
Risk of injury (burns) to personnel who touch the device, depending on the environmental and
operating conditions.
Obey fully the local occupational health and safety regulations. Only approved personnel can
do work on electrical components of the measuring device.