Technical data – KROHNE OPTIMASS 7000 EN User Manual
Page 38
6
TECHNICAL DATA
38
OPTIMASS 7000
www.krohne.com
04/2013 - 4001016302 - MA OPTIMASS 7000 R02 en
Maximum end loadings
• These (axial) loads have been calculated, based on 316L schedule 40 process pipework,
where un-radiographed butt welds have been used in pipe joints.
• The loads shown are the maximum permitted static load. If loads are cycling (between
tension and compression) these loads should be reduced. For advice, consult the
manufacturer.
OPTIMASS 7300C- Aluminium converter
housing - heating jacket / insulation
40
55
T6
T80
70
T5
T95
100
T4
T125
145
T3 - T1
T170
50
70
T4
T95
100
T3 - T1
T125
60
60
T4 - T1
T85
65
1
65
T4 - T1
T90
OPTIMASS 7300C - Stainless Steel
converter housing - no heating jacket /
insulation
40
55
T6
T80
75
T5
T95
120
T4
T130
150
T3 - T1
T160
50
75
T5
T95
115
T4
T130
135
T3 - T1
T145
55
55
T4 - T1
T80
OPTIMASS 7300C - Stainless Steel
converter housing - heating jacket /
insulation
40
55
T6
T80
70
T5
T95
100
T4
T125
145
T3 - T1
T170
50
70
T5
T95
75
T4 - T1
T100
55
55
T4 - T1
T80
1 depending on I/O option. Please call for more information.
Size
06
10
15
25
40
50
80
Titanium
Titanium
Titanium
Titanium
Flanges
19kN
25kN
38kN
60kN
80kN
170kN
230kN
Hygienic (all connections)
1.5kN
2kN
5kN
9kN
12kN
12kN
30kN
Stainless Steel / Hastelloy
Stainless Steel / Hastelloy
Stainless Steel / Hastelloy
Stainless Steel / Hastelloy® / Tantalum
/ Tantalum
/ Tantalum
/ Tantalum
Flanges
19kN
25kN
38kN
60kN
80kN
80kN
170kN
Hygienic (all connections)
1.5kN
2kN
5kN
9kN
12kN
12kN
18kN
CAUTION!
The maximum permitted end loading on size 15 meters fitted with ½¨ ASME flanges is 19kN
MA_OPTIMASS_7000_en_042513_4001016302_R02.book Page 38 Tuesday, July 2, 2013 5:23 PM