DEVICE DESCRIPTION

14

OPTIMASS 7000

www.krohne.com

04/2013 - 4001016302 - MA OPTIMASS 7000 R02 en

Thermal shock

Thermal shock

Thermal shock

Thermal shock

Thermal shock occurs when there is a sudden and extreme change (shift) in process
temperature. To avoid thermal shock, refer to the following table for the maximum temperature
shift.

Meter

Meter

Meter

Meter

Maximum temperature shift

Maximum temperature shift

Maximum temperature shift

Maximum temperature shift

Titanium

+130°C / +266°F

Stainless Steel

+80°C / +176°F

Hastelloy®

+80°C / +176°F

Tantalum

+80°C / +176°F

CAUTION!
Operation outside these limits may result in shifts in density and mass flow calibration. Repeated
shocking may also lead to premature failure of the meter! However, higher thermal shocks are
possible at lower working pressures. For more information, please contact your nearest
representative.