INSTALLATION

3

25

OPTIMASS 7000

www.krohne.com

04/2013 - 4001016302 - MA OPTIMASS 7000 R02 en

Reference conditions

3.4.10 Purge ports

If the meter has been ordered with a purge port, it will be supplied with NPT female connections
which are clearly marked. The connections are sealed with NPT plugs and PTFE tape.

3.4.11 Zero calibration

The procedure for zero calibration is contained in the converter handbook. However, the
following information should be considered when installing the meter.

Temperature [°C / °F] 1

Time [minutes] for Stainless Steel, Hastelloy® and Tantalum

measuring tubes

10

15

25

40

50

80

40 / 104

105

90

60 / 140

190

240

80 / 176

330

480

100 / 212

495

800

1 Measured at the centre of the measuring tube.

Ambient temperature

+25°C / +80°F

Heating medium

Hot liquid

Heating medium temperature

Titanium

+150°C / +302°F

Stainless Steel, Hastelloy® and Tantalum

+100°C / +212°F

CAUTION!
DO NOT remove these plugs.

The meter is factory sealed with a dry nitrogen gas fill and if moisture is allowed to enter the
meter casing it will cause damage. The plugs should only be removed to purge the meter casing
in the event that the primary measuring tube fails.

If it is suspected that the primary measuring tube has failed, de-pressurise the meter and
remove it from service, as soon as it is safe to do so.