Technical data – KROHNE OPTIFLUX 7300 EN User Manual

Page 32

background image

6

TECHNICAL DATA

32

OPTIFLUX 7300

www.krohne.com

01/2011 - 4001163301 - MA OPTIFLUX 7300 R01 en

Low flow cut-off

Low flow cut-off

Low flow cut-off

Low flow cut-off

Function

Switching point and hysteresis separately adjustable for each output, counter

and the display.

Switching point

Set in increments of 0.1.

0....20% (current output, frequency output) or

0...±9.999 m/s (pulse output)

Hysteresis

Set in increments of 0.1.

0...5% (current output, frequency output) or 0...5 m/s (pulse output)

Time constant

Time constant

Time constant

Time constant

Function

The time constant corresponds to the elapsed time until 67% of the end value

has been reached according to a step function.

Settings

Set in increments of 0.1.

0...100 s

Status output / limit switch

Status output / limit switch

Status output / limit switch

Status output / limit switch

Functions and settings

Adjustable as automatic measuring range conversion, display of flow

direction, counter overflow, error, switching point or empty pipe detection.

Valve control with activated dosing function.

Status and/or control: ON or OFF

Operating data

Basic I/Os

Basic I/Os

Basic I/Os

Basic I/Os

Modular I/Os

Modular I/Os

Modular I/Os

Modular I/Os

Ex i

Ex i

Ex i

Ex i

Active

-

U

int

= 24 VDC

I ≤ 20 mA

open:

I ≤ 0.05 mA

closed:

U

0, nom

= 24 V

at I = 20 mA

-

Passive

U

ext

≤ 32 VDC

I ≤ 100 mA

R

L, max

= 47 kΩ

R

L, min

= (U

ext

- U

0

) / I

max

open:

I ≤ 0.05 mA at U

ext

= 32 VDC

closed:

U

0, max

= 0.2 Vat I ≤ 10 mA

U

0, max

= 2 Vat I ≤ 100 mA

-

NAMUR

-

Passive to

EN 60947-5-6

open:

I

nom

= 0.6 mA

closed:

I

nom

= 3.8 mA

Passive to

EN 60947-5-6

open:

I

nom

= 0.43 mA

closed:

I

nom

= 4.5 mA

U

i

= 30 V

I

i

= 100 mA

P

i

= 1 W

C

i

=10 nF

L

i

= 0 mH