beautypg.com

Installation – KROHNE OPTIFLUX 7300 EN User Manual

Page 12

background image

3

INSTALLATION

12

OPTIFLUX 7300

www.krohne.com

01/2011 - 4001163301 - MA OPTIFLUX 7300 R01 en

3.4.3 Flange deviation

3.4.4 T-section

3.4.5 Vibration

CAUTION!
Max. permissible deviation of pipe flange faces:
L

max

- L

min

≤ 0.5 mm / 0.02"

Figure 3-4: Flange deviation

1 L

max

2 L

min

Figure 3-5: Distance after T-sections

1 ≥ 10 DN

Figure 3-6: Avoid vibrations