TECHNICAL DATA

24

OPTIFLUX 5000

www.krohne.com

10/2011 - 4000686202 - MA OPTIFLUX 5000 SW R02 en

6.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

Faraday's law

Application range

Electrically conductive fluids

Measured value

Measured value

Measured value

Measured value

Primary measured value

Flow velocity

Secondary measured

value

Volume flow, mass flow, electrical conductivity, coil temperature

Design

Features

Sandwich version with optimized flow tube

Modular construction

The measurement system consists of a flow sensor and a signal converter.

It is available as compact and as separate version. More information about

the signal converter can be found in the documentation of the signal

converter.

Compact version

With IFC 100 converter: OPTIFLUX 5100 C

With IFC 300 converter: OPTIFLUX 5300 C

Remote version

In wall (W) mount version with IFC 100 converter: OPTIFLUX 5100 W

In field (F), wall (W) or rack (R) mount version with IFC 300 converter:

OPTIFLUX 5300 F, W or R

Nominal diameter

DN2.5...100 / 1/10...4"

Measurement range

-12...+12 m/s / -40...+40 ft/s

Measuring accuracy

Reference conditions

Medium: water

Temperature: 20°C / 68°F

Inlet section: 10 DN

Outlet section: 5 DN

Flow velocity: > 1 m/s / > 3 ft/s

Operating pressure: 1 bar / 14.5 psig

Valve closing time variation: < 1 ms

Wet calibrated on EN 17025 accredited calibration rig by direct volume

comparison.

Maximum measuring

error

Related to volume flow (MV = Measured Value).

These values are related to the pulse / frequency output.

The additional typical measuring deviation for the current output is ±10 μA.

For detailed information refer to

Measuring accuracy on page 29.

Repeatability

±

0.1% of MV, minimum 1 mm/s

Long term stability

±

0.1% of MV

Special calibration

On request