3.4 Pre-installation requirements

Make sure that you have all necessary tools available:

• Allen key (4 mm)
• Small screwdriver
• Wrench for cable glands
• Wrench for wall mounting bracket (remote version only)
• Torque wrench for installing flowmeter in pipeline

3.5 Installation conditions

3.5.1 Inlet and outlet

3.5.2 Mounting position

Figure 3-2: Recommended inlet and outlet

1 ≥ 5 DN
2 ≥ 2 DN

Figure 3-3: Mounting position