OPERATION

4

35

OPTISENS PH 9500

www.krohne.com

08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en

• Press ^ to confirm.

i

The slope of the sensor is displayed (should be near 59 mV in an intact sensor).

• Press ^ to confirm.

i

The message store cal. value? is displayed on the screen. The converter asks if the new
calibration values should be stored.

• Choose yes to store the calibration values. Choose no to discard the results.
• Press ^ to confirm.

i

Prior returning to the measuring display, you are asked if the configuration should be
stored.

• Choose yes using or to store the new calibration values.

i

You have completed the pH calibration.

• If you want to return to the measuring mode, press ^ several times until you reach this mode.

Step 5: re-installing the sensor

• After the calibration procedure, rinse the sensor with tap water.
• Reinstall the sensor into its assembly, refer to

Installing the sensor

on page 17.

Step 6: switching back to measurement

• Deactivate the function "manual hold" again.

INFORMATION!
If an error occurs during the calibration procedure, the display shows an error message.
Possible causes for an error are:

Slope too flat.

Wrong buffer solution.

Wrong electrical connection.

Moisture in the sensor connector.