beautypg.com

Operation – KROHNE OPTISENS PH 9500 User Manual

Page 29

background image

OPERATION

4

29

OPTISENS PH 9500

www.krohne.com

08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en

Step 2: Configure /adjust the temperature sensor

• If necessary, enter the temperature correction in Kelvin so that the signal converter shows the

same temperature as the reference thermometer. Press ^ to confirm the entered value. The
temperature sensor has been adjusted.

Step 2a: probe Pt100/1000

Read the currently measured temperature of the Pt100/1000 temperature sensor from the measurement

screen and write it down.

Measure the temperature with a reference thermometer and check if it deviates from the temperature

measured by the Pt100/1000.

Press > for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu setup is highlighted.

Press or until the main menu setup is highlighted.

MAIN MENU

A quick setup

B test

> C setup

D service

Press > to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,

beneath the submenu process input A is highlighted. Choose process input A or B where ever

pH is configurated.

Press or to select process input A or process input B.

Press > to enter the chosen menu process input A/B.

You are on the second submenu level. Press or until the submenu temperature is

highlighted.

Press > to enter the chosen menu. The submenu probe is highlighted.

Press or until the submenu Pt100/1000 is highlighted.

Press ^ to enter the chosen menu.