OPERATION

4

27

OPTISENS PH 9500

www.krohne.com

08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en

4.3 Calibration

4.3.1 Temperature compensation

There are three basic options for temperature compensation:

• automatic: the signal converter will automatically compensate temperature influences using

the information of a Pt100 or Pt1000 temperature sensor.

• manual: the signal converter will compensate temperature influences using a manually

entered value; this option only makes sense if the temperature of the measured medium is
quite constant.

• off: temperature compensation is disabled.

The menu for the type of temperature compensation offers the following options:

• Pt1000: choose this option if there is an external Pt1000 temperature sensor connected to the

signal converter.

• Pt100: choose this option if the sensor has an integrated Pt100 temperature measurement or

if there is an external Pt100 temperature sensor connected to the signal converter.

• manual: choose this option if there is no integrated or external temperature sensor

connected to the signal converter.

• process input A: choose this option if there is a temperature sensor connected to input A of

the signal converter.

INFORMATION!
If you choose no compensation, the measured pH value will most likely deviate from the actual
pH value. The reason is that the pH value of a specific medium varies depending on the
temperature of the medium.

INFORMATION!
In the basic configuration the setting of the temperature compensation is Pt100. If you use a pH
sensor without Pt100/1000 than select the temperature compensation "manual" and enter the
value. Otherwise an error will be displayed.