Operation – KROHNE OPTISENS PH 9500 User Manual

Page 26

background image

4

OPERATION

26

OPTISENS PH 9500

www.krohne.com

08/2012 - 4001929601 MA OPTISENS PH 9500 R01 en

C1.15.5

C2.15.5

temp. comp.

Menu for activating the temperature compensation parameters for the

measurement.

Options:

Linear: linear temperature compensation.

Off: temperature compensation is disabled.

C1.16

C2.16

ORP cal.

Menu item for calibrating the ORP sensor.

C1.16.1

C2.16.1

prepare calibration

View actual ORP value.

C1.16.2

C2.16.2

ref. value ORP

Enter the reference value of the probe in mV.

C1.16.3

C2.16.3

start calibration?

Start calibration procedure.

C1.16.4

C2.16.4

stored value

View stored value of calibration.

C1.17

C2.17

pH cal.

Menu item for calibration the pH sensor.

C1.17.1

C2.17.1

temp.comp

Menu for activating the temperature compensation parameters for the

calibration.

Options:

off: temperature measurement is disabled.

manual: temperature value has to be entered manually.

automatic: temperature measurement is performed as configured.

C1.17.3

C2.17.3

calib. buffer 1

Start calibration procedure in buffer 1.

C1.17.4

C2.17.4

pH 1 set value

Enter the reference value of the probe in pH.

C1.17.6

C2.17.6

calib. buffer 2

Start calibration procedure in buffer 2.

C1.17.7

C2.17.7

pH 2 set value

Enter the reference value of the probe in pH.

CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.

Level

Designation / function

Settings / descriptions