8.2 NAMUR NE 107 status signals

8.3 Supported Modbus function codes

Function Code 0x01: Read Coils

Function Code 0x02: Read Discrete Inputs

F

Failure

Failure

Failure

Failure
Output signal invalid due to malfunction in the signal converter.

C

Function Check

Function Check

Function Check

Function Check
Output signal temporarily invalid due to ongoing work on the signal converter.

S

Out of specification

Out of specification

Out of specification

Out of specification

• Deviations from the permissible ambient or process conditions determined by

the signal converter itself through self-monitoring.

• Faults in the signal converter itself indicate that the measuring uncertainty of

flow sensors or deviations from the set value in actuators is probably greater
than expected under operating conditions.

M

Maintenance required

Maintenance required

Maintenance required

Maintenance required
Although the output signal is valid, the wear reserve is nearly exhausted or a
function will soon be restricted due to operational conditions.

Request

Response

Error

Function

0x01

Function

0x01

Function

0x81

Starting Address

Hi

0x00 to 0xFF

Byte Count

n

Exception Code

0x01 / 0x02 /

0x03 / 0x04

Starting Address

Lo

0x00 to 0xFF

Coil n Status

Quantity of Coils

Hi

n (0x00 to 0x07)

...

...

Quantity of Coils

Lo

n (0x01 to 0xFF)

max Quantity:

0x07D0

Request

Response

Error

Function

0x02

Function

0x02

Function

0x82

Starting Address

Hi

0x00 to 0xFF

Byte Count

n

Exception Code

0x01 / 0x02 /

0x03 / 0x04

Starting Address

Lo

0x00 to 0xFF

Input n Status

Quantity of

Registers Hi

n (0x00 to 0x07)

...

...

Quantity of

Registers Lo

n (0x01 to 0xFF)

max Quantity:

0x07D0