Modbus protocol – KROHNE MFC 400 Modbus User Manual
Page 25
MODBUS PROTOCOL
5
25
MFC 400
www.krohne.com
05/2013 - 4002525102 - AD Modbus MFC 400 R02 en
5.8.14 NAMUR NE 107 Variable Event Group(s)
The following eight event groups can be mapped to any status signal.
For further information refer to section "Status messages and diagnostic information" in the
signal converter standard manual.
Legal values for those registers:
• 128: Failure (F)
• 32: Out Of Specification (S)
• 16: Function Check (C)
• 4: Maintenance Request (M)
• 1: Information (I)
Modbus Function Codes "Read Holding Registers" (0x03) and "Write Multiple Registers" (0x10)
Holding
Register
Description
Status Group
Type
No. of
registers
Default Value
Display
Fct. No.
52016
52016
52016
52016
(0xCB30)
Status Signal of
Event Group 15
Proc: Current Input
Byte
1
128
Failure (F)
C1.5.9
52017
52017
52017
52017
(0xCB31)
Status Signal of
Event Group 14
Electr: IO Connection
Byte
1
32
Out Of
Specification (S)
C1.5.14
52018
52018
52018
52018
(0xCB32)
Status Signal of
Event Group 13
Proc: Signal Search
Byte
1
128
Failure (F)
C1.5.8
52019
52019
52019
52019
(0xCB33)
Status Signal of
Event Group 12
Proc: 2 Phase Flow
Byte
1
32
Out Of
Specification (S)
C1.5.10
52020
52020
52020
52020
(0xCB34)
Status Signal of
Event Group 11
Proc: Signal Low
Byte
1
32
Out Of
Specification (S)
C1.5.7
52021
52021
52021
52021
(0xCB35)
Status Signal of
Event Group 10
Config: Totaliser
Byte
1
32
Out Of
Specification (S)
C1.5.12
52022
52022
52022
52022
(0xCB36)
Status Signal of
Event Group 9
Proc: System Control
Byte
1
1
Information (I)
C1.5.11
52023
52023
52023
52023
(0xCB37)
Status Signal of
Event Group 8
Electr: Power Failure
Byte
1
32
Out Of
Specification (S)
C1.5.13