Modbus protocol, 2 device status – KROHNE MFC 400 Modbus User Manual
Page 15
MODBUS PROTOCOL
5
15
MFC 400
www.krohne.com
05/2013 - 4002525102 - AD Modbus MFC 400 R02 en
5.8.2 Device Status
Modbus Function Codes "Read Discrete Inputs" (0x02)
Discrete
Inputs
Name
Description
Type
No. of
registers
Values
10000
10000
10000
10000
(0x2710)
Status of Custody
Transfer Lock
Indicates if custody transfer lock
is active or no
Bit
1
0 = unlocked
1 = locked
10001
10001
10001
10001
(0x2711)
Are Changes
Made?
Indicates if there are unsaved
changes. "Apply Changes" to
save them, "Discard Changes" to
work with previous settings
Bit
1
0 = no changes made
1 = unsaved changes
detected
Modbus Function Codes "Read Input Registers" (0x04)
Input
Register
Name
Type
No. of
registers
Values / Units
Display
Fct. No.
39000
39000
39000
39000
(0x9858)
Actual Operation Mode
Long
2
1 = Stop
2 = Startup
3 = Measuring
5 = Standby
B2.16
39002
39002
39002
39002
(0x985A)
Device Operation Time
Float
2
[s]
B2.1
39004
39004
39004
39004
(0x985C)
NE 107 Status Groups
Long
2
For further information refer to
NAMUR NE 107 Event Group(s)
on page 32.
-
39100
39100
39100
39100
(0x98BC)
NE 107 Device Status
Byte
1
Bit 7 = Failure (F)
Bit 6 = reserved
Bit 5 = Out of Specification (S)
Bit 4 = Function Check (C)
Bit 3 = reserved
Bit 2 = Maintenance required (M)
Bit 1 = reserved
Bit 0 = Information (I)
-