Modbus protocol, 1 device control – KROHNE MFC 400 Modbus User Manual
Page 14
5
MODBUS PROTOCOL
14
MFC 400
www.krohne.com
05/2013 - 4002525102 - AD Modbus MFC 400 R02 en
5.8.1 Device Control
The "Device Control" offers some basic functionality to operate with the signal converter.
Therefore, the Modbus interface provides five coils that can be accessed via Modbus Function
Code "Write Single Coil" (0x05).
Write a coil to value 1
value 1
value 1
value 1 (ON) to initiate the action.
The flow sensor can be switched between three modes. Use the following Modbus register to
request a change of the operation mode. The actual operation mode can be read via Modbus
register 39000 (details on page 15).
Modbus Function Code "Write Single Coil" (0x05)
Coil
Address
Name
Description
Display
Fct. No.
1000
1000
1000
1000
(0x03E8)
Restart Device
Restart entire signal converter
D2.2.1
1001
1001
1001
1001
(0x03E9)
Reset Errors
Clears the system error flags
A3.1
1002
1002
1002
1002
(0x03EA)
Apply Changes
Apply latest changes of configuration
-
1003
1003
1003
1003
(0x03EB)
Discard Changes
Discard all of the configuration changes made since the last "Apply
Changes"
-
1004
1004
1004
1004
(0x03EC)
Reset to Factory
Data
Resets the signal converter to factory configuration
C6.6.3
Modbus Function Codes "Read Holding Registers" (0x03) and "Write Multiple Registers" (0x10)
Holding
Register
Name
Description
Type
No. of
registers
Values
Display
Fct. No.
51000
51000
51000
51000
(0XC738)
Operation
Mode
Set the actual operation mode
of the flow sensor
Byte
1
1 = Stop Mode
3 = Measure Mode
5 = Standby Mode
A.9